Software Release Arc_1201; New Features And Solved Issues Of Software Arc_1201 - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2

Software Release ARC_1201

Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
* On the Agfa HealthCare Library only the latest software release can be downloaded.
2.1

New Features and solved Issues of Software ARC_1201

A new Error Viewer version 2.0.9.0 is released together with software ARC_1201.
New features of the Error Viewer:
o Connection between digitizer and PC is checked periodically: Status indicator
switches to "red triple flashing" if a "ping" to the configured PC fails and to
"fast red blinking" if the connection to the Error Viewer fails. The network
connection failure has the higher priority.
o Title of the Error Viewer panel is changed to "AGFA Digitizer Remote Display".
o If an error is displayed, the user can browse through all previous errors i.e.
errors messages which were closed by pressing the "OK" button.
o Displays now the status "connected" or "offline" in the new status bar.
o The digitizer checks now the version of the Error Viewer: If this version is not
compatible to the digitizer software or has severe known bugs, the digitizer
switches the status indicator to "red fast blinking" to indicate that the
ErrorViewer must be updated to the version which is integrated in the digitizer
software.
Reliability of the barcode reading improved.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
04-2013
05/2012
Software ARC_1201 is installed in digitizers with SN ≥
1039 and additionally in SN 1014, 1016, 1018, 1019,
1020, 1028 ex factory.
The NX workstation connected to the digitizer requires
software ≥ NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
DD+DIS033.12E, Service Bulletin No. 01,
Digitizer CR 10-X released for controlled Delivery
(Intranet Link
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 5 of 13
Type 5151 / 100 / 110 / 200 / 210
Software Releases and Patches
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Solution Structure Data Document, Document ID*
30781759
(Intranet Link
CR/DR Interoperability Matrix, Document ID*
31333326
(Intranet Link
/
Extranet
Link)
Repair and Service
/
Extranet
Link)
/
Extranet
Link)
Agfa Company Confidential