Safety Notes For Service Activities; General Safety Notes For All Service Activities; Safety Notes For Installation Planning - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
3

Safety Notes for Service Activities

3.1

General Safety Notes for all Service Activities

CAUTION:
Sharp edges inside the digitizer: Cut or abrasion possible.
Be careful at any maintenance work.
CAUTION:
Image plate or cassette unit can be damaged.
Do not insert and pull out cassette as long as the digitizer is switched off.
WARNING:
120 mW Laser (Class 3B) in the Optic Module may cause eye injuries.
Avoid direct and indirect eye contact.
Do not keep tools in the laser beam when the device is switched on.
Do not open the cover of the Optic Module.
3.2

Safety Notes for Installation Planning

Warning:
Images can be lost due to power failure.
Connect the equipment to an un-interruptible power supply (UPS) or an institutional
standby generator.
WARNING:
When different combinations of equipment are used in various medical
environments a potential difference (V) can exist between the protective earths
in different localities. If the protective earthing fails this potential difference can
cause a HAZARD for the OPERATOR or for the PATIENT.
To comply with ISO 60601-1 (annex I) all computers and peripherals must be
connected to the same power source.
Always connect the associated monitor to the same Uninterruptible Power
Source (UPS) as the PC.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.1 / Page 6 of 11