Scanning The Image Plate - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2.3

Scanning the Image Plate

(1)
On the NX workstation log-off the current user and log-in as "crservice".
(2)
Open the Internet Explorer.
(3)
Enter the digitizer IP address in the address window. See also NOTE below.
NOTE:
If the digitizer IP address is not known, look it up in the NX configuration tool:
Double-click the icon <Start NX Configurator> on the desktop.
Select: Load active configuration
Go to: Devices > Digitizer Configuration
This page shows the digitizer IP address.
Quit the NX Configurator (Close without saving any changes).
(4)
Enter username and password (obtained during the web based training).
(5)
Select the service menu: <Shading Calibration>
(6)
Select calibrate.
(7)
Press <confirm> in the Error
Viewer, when the Service user
interface shows following
message: Please confirm at
system to continue.
Waiting for confirm ...
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.6 / Page 7 of 10
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Adjustments and Calibrations
Figure 2
Figure 3
Agfa Company Confidential