5.4

Verifying the Installation

NOTE:
Verifying the installation means: Expose a flatfield and evaluate the image quality.
Following items are verified with the check:
The configuration was successful: Images arrive at the workstation.
Image quality is OK.
(1)
Erase a cassette.
(2)
Use the erased cassette to
expose a flatfield.
 The black cassette side
must be facing up.
 The collimated area
should be larger than the
cassette.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
03-2013
Controls, Connections, and Setup Procedures
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
A
Press the "Erase" button.
Insert cassette with
B
black side facing up.
C
Wait approx. 3 minutes.
D
Remove cassette.
Figure 61
1. Insert CU Filter
2. Expose Cassette:
3. Rotate cassette by 180°
4. Expose again
1,3 m
5. Remove CU Filter
Collimated
Area
Black side
facing up
Figure 62
Agfa Company Confidential
install_poster_12.cdr
Large Focus
12 mAs
75 kVp
install_poster_013.cdr
Chapter 1 / Page 43 of 47