Image Faults vertical to Image Plate Run
7.1
Banding
Banding stripes are visible in the clinical
Symptom
image or a flatfield image.
Inhomogeneous movement of the slow scan drive.
Cause
(1)
Check the table where digitizer is positioned. It must be a fixed position. No
Solution
vibrations may disturb the scanning.
(2)
If no vibrations from the table are transferred to the digitizer:
Escalate to the next support level.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Figure 5: Example for banding