Selftest error of 'Cassette Detection Light Sensor'.
Meaning: Selftest detects 'Cassette Detection Light Sensor' disconnected.
Reason 1: The cable is defective or disconnected.
Cure 1: Check cable and connectors to 'Cassette Detection Light Sensor'.
Replace 'Y-Cable Power Board to Drawer/Cass. LS'.
Reason 2: 'Cassette Detection Light Sensor' defective.
Cure 2: Replace 'Cassette Detection Light Sensor'.
Reason 3: 'Power Board' defective.
Cure 3: Replace 'Power Board'.
Selftest error of 'Cassette Detection Light Sensor' and 'IP Drawer Light Sensor'.
Meaning: Selftest detects 'Cassette Detection Light Sensor' and
'IP Drawer Light Sensor' disconnected.
Both sensors are connected via one Y-cable.
Reason 1: The cable is defective or disconnected.
Cure 1: Check cables and connectors to
'Cassette Detection Light Sensor' and 'IP Drawer Light Sensor'.
Replace 'Y-Cable Power Board to Drawer/Cass. LS'.
Reason 2: 'Cassette Detection Light Sensor' defective.
Cure 2: Replace 'Cassette Detection Light Sensor'.
Reason 3: 'IP Drawer Light Sensor' defective.
Cure 3: Replace 'IP Drawer Light Sensor'.
Reason 4: 'Power Board' defective.
Cure 4: Replace 'Power Board'.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Message
Chapter 3.3 / 30 of 46
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents