Appendix: Possible Reasons for a failed Shading Calibration

There are following main reasons for a failed shading calibration:
Particles in the optical path  Use the cleaning brush to clean the optical path.
For instructions refer to the digitizer user manual.
Contamination and/or any defects on the image plate  Clean the image plate
according to the cleaning procedure or use another image plate.
Cassette wrong exposed (too high / low dose)  Check exposure parameters.
Cassette not exposed on its entire area (wrong collimation)  Ensure the
cassette is entirely exposed.
PMT exposed to intensive light before calibration (in this case PMT has lower
sensitivity)  wait until the PMT has back the normal sensitivity (time depends
on duration and brightness of the light exposure).