Overview of Adjustments and Calibrations
The following tables give an overview of adjustments and calibrations for the digitizer.
Adjustment Type
None
Calibration Type
Shading calibration
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Use Case
Not applicable: After spare parts exchange no
adjustment is required.
Use Case
After replacement of the PMT with Light
Collector
After replacement of the Optic Module
Whenever stripes in image plate run direction
are visible which cannot be removed by
cleaning with the cleaning brush