Photomultiplier flash memory errors are repaired automatically: The software
doesn't crash anymore if the Photomultiplier flash memory is corrupted.
Galvo adjustment is optimized: The adjustment is checked and repeated once per
minute if required.
Infocounter values like installation/manufacture date and machine running time are
corrected.
Image is not lost anymore for following special case:
- Network is disconnected during scan and
- Cassette is not removed after scan and
- Digitizer is switched off afterwards
2.2

Known Software Issues of Software ARC_1201

When running service menu "Endurance run" without scanning or without erasing: The
status is not updated. It remains on "feeding IP".