Reset Relative Counters; Clear Image Queue - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2.5.6

Reset Relative Counters

With the menu "Reset Relative Counters" the relative counters* of the info counter
Purpose
are cleared.
This may be required to easily track new info counter entries after a service
intervention.
* Number of test cycles per format; number of start-ups and resets; Retries; Incorrect User Handling;
Cassette and IP jams; Error List; Warning List
(1)
Select service menu: <Reset
Usage
Relative Counters>
(2)
Click: <OK>
2.5.7

Clear Image Queue

There are cases, where an image cannot be transmitted, and stays in the queue
Purpose
forever, despite the processing station software is running. This may happen during
setup of the digitizer. Examples:
An image is scanned while the network is disconnected. At the same time the
destination settings are changed (e.g. IP address of NX).
Destination AE title is wrong configured in the digitizer.
In this case after each subsequent scan a "destination not available" or "destination is
invalid" warning is displayed.
With the menu "Clear Image Queue" all images in the image queue can be erased.
WARNING:
By clearing the image queue all images in the queue of the digitizer are erased.
Retake possible.
Before clearing the image queue, check all other troubleshooting options to get the
image transmitted to the destination.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.2 / Page 25 of 31
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Tools and auxiliary Means
Figure 22
Agfa Company Confidential

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents