2.3. PATIENT POSITIONING

CAUTION: Before any table movement, overall tilting or tabletop movements, make sure there are
no obstacles that can collide during movement.
WARNING
Controls: to ensure patient safety, it is essential that the operator keeps eye contact with the patient
whereas he controls table motion.
2.3.1.

ACCESSORIES INSTALLATION

 In order to ensure equipment reliability, use only accessories supplied with the table (1 footrest,
1 stool, 2 handles, 2 winches and compression band). The user is held solely responsible if use
of accessories not supplied by the manufacturer.
 The handles can be placed anywhere in the rails of the tabletop, except at the center of rails
where a slot is. Flip the lever on the side opposite to the handle, insert the handle into rail to
the desired location and then flip the lever on the other side to lock the handle
 The compression winches can be placed anywhere in the rails of the tabletop: put the table in
horizontal position, loosen the small knob, insert the winch support by one or the other end of
the rail, place the winch to the desired location and tighten this knob. For the second winch,
repeat the operation and place it symmetrically to the first one. Then, put each metal end of
band inside winch slot and make compression by rotating assymetrically axes of winches
thanks to large knobs.
 The footrest can be placed anywhere in the rails of the tabletop, except at the center of rails
where a slot is: put the table in horizontal position, grasp the footrest with two handles, tighten
the left one and insert the white wheels precisely aligned in the rails of the tabletop. Slide the
footrest to the desired position and release the left handle that blocks the brake. To remove the
footrest, grasp the two handles, tighten the left one and press the tab on the right. Slide the
footrest to get completely out of rails.
 The footrest can not be positioned at centre of tabletop
2.3.2.

PATIENT INSTALLATION

 If accessories are to be used to secure the patient in position, refer first to the relevant fitting
instructions.
 Set the table in horizontal (
(tilting control).
 In the case of recumbent patients, it is advised to move the tabletop fully towards the head end
an assistant when transferring a patient from a trolley.
Digital Remote Controlled Table with dynamic flat panel detector
and opposite to the column
User's Manual
or
&
) or vertical
, and to lower the table
23
position according to the examination