Faults With Error Codes - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
2

Faults with Error Codes

The digitizer Error Viewer on the NX workstation displays an error code,
Symptom
e.g. ARC1001 (25613).
NOTE:
Meaning of the error codes ARCXXXX (YYYYY):
The error code ARCXXXX is the service message number.
The error code (YYYYY) is the R&D internal error code.
The digitizer software detected a malfunction of the digitizer.
Cause
Solve the problem as described in the service description of the error code.
Solution
To look-up a digitizer error code (e.g. ARC1001) proceed as follows:
A) Look-up the error "online" (Access to NX Workstation or PC in the network
available):
(1)
On the NX Workstation or network PC start the internet browser.
(2)
Enter the IP address of the digitizer in the address window.
(3)
Login with username and password provided in the web based training.
(4)
Look-up the error (e.g. ARC1001) in menu "Explain Error Code".
B) Look-up the error "offline" (No access to NX Workstation):
(1)
Download the latest digitizer software file from the
Agfa HealthCare Library.
(2)
Unzip the software file.
(3)
Unzip the ARC_XXXX.zip* file.
(4)
Using Excel, look-up the error in file "errors.csv".
C) Look-up the error in section 9 of this document.
*
XXXX represents the software version
NOTE:
The R&D internal error codes YYYYY can also be looked up in menu "Explain
Error Code" or the file "errors.csv". The error descriptions are however not
prepared for the service user.
With each error a file "malfunction file" with the syntax
MALF_<Type Nr.>_<SN>_<Error Code>_<Date>_<Time>.zip
(e.g. MALF_5151_01050_5145_20Aug2013_08-29-17.zip) is created on the
digitizer USB flash drive. This file contains all configuration and logfiles, which are
required for fault analysis when escalating a problem. All available malfunction files
are part of the backup ZIP file. Malfunction files are available as of software
ARC_1404.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
.
Agfa Company Confidential
Troubleshooting
Chapter 3.3 / 5 of 46

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents