DD+DIS155.11E
6.2
Dust Stripes
Dust stripes are visible in the clinical
Symptom
image or a flatfield image.
Dirt or dust in the scan path.
Cause
(1)
Use the cleaning brush to clean the mirror and the entrance surface of the light
Solution
collector by moving the brush a few times. For instructions refer to the digitizer
user manual.
(2)
Expose and scan a flatfield and evaluate again.
For instructions how to create a flatfield image refer to chapter 3.6, Adjustments
and Calibrations
(3)
If unacceptable image flaws suntil occur: Escalate to the next support level.