Boot-Up Of The Digitizer - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
4

Boot-up of the Digitizer

The following flowchart illustrates the boot-up of the digitizer:
Approximate
time in
seconds
Digitizer
0
Switch-On
Boot-up of PMI and
17
Power Board finished
Load data and
check E-labels
Send Self Test
40
Signal to all SW modules
Selftest okay?
IP Clean-up
43
Cycle
Cassette or IP
found?
NO
60
Adjust galvo
Set drawer motor
100*
to HOME
Initialization okay?
NX Viewer
running?
105*
Digitizer idle
* time depends on galvo adjustment time. Can be up to 5 minutes.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Digitizer Default Boot-Up Sequence
Status LED starts triple red blinking
if no network connection to NX.
Status LED starts
if no connection to Error Viewer.
Error message
in Error Viewer.
NO
Boot-up stops
YES
Warning message
in Error Viewer
YES
Rescue Cycle: Digitizer
tries to solve problem.
Error message
in Error Viewer.
NO
Boot-up stops
YES
Wait till the Digitizer
NO
receives "ready for
identification" message
YES
515102bb.cdr
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 2 / Page 12 of 19
Type 5151 / 100 / 110 / 200 / 210
Functional Description
fast red blinking
Figure 10
Visible / a
udible
1 sec.
Effects
Message
"Digitizer start-up"
(4198) in Error Viewer
Drawer Motor
briefly
audible
Agfa Company Confidential