DD+DIS170.11E
8.8
Interface Data
Interface to NX workstation
* Gigabit Ethernet is not supported.
8.9
Service Data
Info counter
Error diagnostics built in
Remote / preventive service capable
8.10
Functional Data Image Plates
Image plate type
Image retention
Image plate Usage
9
Safety Standards
The digitizer complies with:
EMC*
Safety
* EMC = Electromagnetic Compatibility
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Within 1 hour
After 2 hours
After 24 hours
North American Standards
Federal Communications Commission
Canadian Standards Association
Underwriters Laboratories
Code of Federal Regulations
CSA 22.2 No. 601.1.2
FCC part 15
UL60601-1:2003
CSA C 22.2 No. 601.1:1990 +
S1:1994 + A2:1998
CFR parts 1040.10 and
1040.11
CSA-E60825-1-03
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 21 of 24
Type 5151 / 100 / 110 / 200 / 210
Installation Planning
Via Ethernet 10/100 Base Ethernet
(auto-negotiate)*
Yes
Yes
Yes
CR MD1.0
Recommended readout time
70 % of the stored energy is still present with
no visible loss of information upon readout.
Image retention still exceeds 45 %.
Slightly reduced image quality
Image plate expires after 5 years
European Norm and International
Standards
European Norm
International Electrotechnical Commission
EN 60601-1-2:2007
EN ISO 13485: 2003
IEC 60601-1:1988 + A1:1991 +
A2:1995
IEC 60601-1: 2005
IEC 60825-1: 2007
Agfa Company Confidential
1
2
3