Connection Problems - AGFA CR Reader User Manual

Hide thumbs Also See for CR Reader:
Table of Contents

Advertisement

CR Reader, CR Advanced Reader, CR Multiformat Reader | Operating CR Reader, CR Advanced

Connection Problems

In case the status indicator of the digitizer is blinking red, the user should look
at the "status" of the Digitizer Remote Display to decide whether digitizer
internal problems or connection problems occurred.
If an error message is displayed on the NX PC, the user is informed which
actions to perform to solve the problem.
In case no error message is displayed on the screen, a connection problem
occurred.
Condition
Connection
problem be-
tween digitizer
and the Digitiz-
er Remote Dis-
play.
Connection
problem be-
tween digitizer
and NX PC.
CAUTION:
Operation failure of the device can cause delayed diagnosis.
Check if Digitizer Remote Display is running.
Message at
Digitizer Re-
mote Display
No error mes-
sage on NX PC.
Reader and CR Multiformat Reader | 65
Status indica-
Action
tor
Red blinking
Check if Digitizer Re-
fast
mote Display is running.
Start/restart Digitizer
Remote Display.
Red blinking -
Check the Ethernet ca-
3 pulses
bles. If the error remains,
restart PC and digitizer
or call service.
2591D EN 20191028 1712