No Error Code - Status Led Triple Red Blinking - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
Error viewer not started up.
Cause 1
Go to Start > All Programs > Startup > Error Viewer to start the Error
Solution 1
Viewer.
Expected Error Viewer version not installed. Each software version has its dedicated
Cause 2
Error Viewer.
Follow the software installation instructions which are part of the software zip file on the
Solution 2
library to:
Uninstall the Error Viewer.
Install the Error Viewer which is part of the digitizer software.
If the digitizer is connected to SE: Prepare the digitizer for usage with the old error
viewer protocol as described in the SE service manual chapter 5 (Configuration) or as
described in the software installation instructions for software ≥ ARC_1404 (for
software upgrades incl. formatting only)
5

No Error Code - Status LED triple red blinking

No error displayed in the Error Viewer on the workstation
Symptom
Status LED is triple red blinking (approx. 3 on/off cycles per second followed
by 1 second off)
Background: Ping sent from digitizer to NX workstation with IP address configured in
digitizer failed.
IMPORTANT:
File ErrorInformation.html with information about the last error and the digitizer
configuration is written to USB flash memory if the connection to the Error Viewer on
the NX workstation is not available (available as of software ARC_1302). This file is
removed again when the Error Viewer is reconnected.
Recommendation for efficient troubleshooting:
(1) Remove the USB flash drive from the digitizer.
(2) Insert it in the NX workstation.
(3) Double-click file: ErrorInformation.html
(4) Follow the instructions.
For an example of file ErrorInformation.html see page 9.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Troubleshooting
Chapter 3.3 / 10 of 46
Agfa Company Confidential