Checking Software Version; Performing Further Activities Depending On Local Regulations; Customer Training - AGFA CR Series Service Manual

Digitizer
Table of Contents

Advertisement

DD+DIS155.11E
4.6

Checking Software Version

(1)
Select service menu
Device Info and check
the digitizer software
version.
(2)
Perform a software update
if a newer software is
available. Refer to the
instructions enclosed to the
software ZIP file.
NOTE:
The latest software is available from the Agfa HealthCare Library:
Computed Radiography => CR Digitizer => CR 10-X / CR 12-X => Software
4.7

Performing further Activities depending on local Regulations

Enclosed a list of possible activities depending on local regulations:
This list is just an example.
#
Activity
1
Electrical check
2
Acceptance Test
(1)
Contact your local service manager concerning the activities which are required
due to local regulations.
4.8

Customer Training

(1)
Hand-over the user manuals.
(2)
Train the customer in following activities:
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 4
03-2013
Controls, Connections, and Setup Procedures
Operating the digitizer, see section 4.8.1.
Usage of plates and cassettes, see section 4.8.2.
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader
Type 5151 / 100 / 110 / 200 / 210
Device Info
Item
Type Number:
Subtype Number:
Serial Number:
Manufacture Date:
Installation Date:
Software Version:
Figure 15
Reference
IEC 62353:2007 Test Documentation on
the Agfa HealthCare Library
Intranet
Link)
Example for Germany: DIN6868
Agfa Company Confidential
Value
5151
100
1001
02.01.2012
15.01.2012
ARC_1013
(Extranet Link
/
Chapter 1 / Page 18 of 47

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents