X-Ray Radiation Conditions

Observe following warning for the installation place of the digitizer and the storage
location of the cassettes with image plates:
WARNING:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this
way, that the annual dose equivalent at the installation place will not exceed 1 mSv.
6

Electrical Connections

Warning:
Images can be lost due to power failure.
Connect the equipment to an un-interruptible power supply (UPS) or an institutional
standby generator.
WARNING:
When different combinations of equipment are used in various medical
environments a potential difference (V) can exist between the protective earths
in different localities. If the protective earthing fails this potential difference can
cause a HAZARD for the OPERATOR or for the PATIENT.
To comply with ISO 60601-1 (annex I) all computers and peripherals must be
connected to the same power source.
Always connect the associated monitor to the same Uninterruptible Power
Source (UPS) as the PC.
WARNING:
Electrical device. Shock possible.
INSTRUCTION:
Position the Agfa products so that it is possible to disconnect the mains power
connection if required.
Local and International wiring regulations must be observed. Check all supplies
and voltages, currents, trips and fuses with the Hospital facilities department or
their engineers.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 11 / Page 14 of 24
Type 5151 / 100 / 110 / 200 / 210
Installation Planning
Agfa Company Confidential
1
2
3

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents