WARNING:
Electrical device. Shock possible.
INSTRUCTION:
Position the Agfa products so that it is possible to disconnect the mains power
connection if required.
Local and International wiring regulations must be observed. Check all supplies
and voltages, currents, trips and fuses with the Hospital facilities department or
their engineers.
WARNING:
The device is a table-top digitizer. The structure and stability of the table used need to
be suitable in relation with the size and weight of the system. Do not use excessive
force when inserting cassettes in the digitizer as the device may slip or drop off the
table. Use a non-slip-mat below the digitizer or other anti-skid measures. The table
should not be subject to excessive shock and vibrations from other sources, as this
may disturb the operation of the digitizer.
WARNING:
Image plate is sensitive for X-rays. Poor image quality possible.
The digitizer and the cassette storage shall be protected against X-ray radiation this
way, that the annual dose equivalent at the installation place will not exceed 1 mSv.
3.3

Safety Notes for Installation

WARNING:
The digitizer is heavy (30 kg; 66 lbs). Risk of injuries when lifting the digitizer.
Use proper foot and hand protection when lifting the digitizer.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 2
03-2013
Safety Guidelines / General Repair Instruction
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.1 / Page 7 of 11
Type 5151 / 100 / 110 / 200 / 210
Repair and Service