Software Release ARC_1302

Release Date
Digitizer SNs
Availability
Prerequisites
Reference Document
* On the Agfa HealthCare Library only the latest software release can be downloaded.
4.1

Solved Software Issues in Software ARC_1302

Solution for Headquarter Complaint HQ_1207130004:
With ARC_1302 also a new version of the digitizer operating system is installed. This
solves the issue with some digitizer USB flash drives, being too slow for software
installation.
In software ARC_1201 a workaround was implemented to prevent a software crash in
case of Photomultiplier flash memory errors. Now the final solution is implemented, to
prevent read and write errors of the Photomultiplier flash memory.
Solved sporadic software issue with error viewer being offline each 15 seconds. The
issue may appear if digitizer and PC are switched on at the same time.
Solved issue with green blinking status LED after scan for several minutes, if an error
occurred during scan.
The feedback messages during service function "endurance run" are now updated with
each step (feeding IP, scanning, erasing etc.). Before, the feedback messages stopped
with "feeding IP", despite endurance run worked fine.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 3
04-2013
09/2012
Software ARC_1302 is installed in digitizers with
SN ≥ 1517 and additionally in SN 1381, 1424, 1428,
1429, 1430, 1433, 1435
The NX workstation connected to the digitizer requires
software ≥ NX 2.0.8700 / 3.0.8700.
For additional system component compatibility
information refer to following documents:
DD+DIS189.12E, Service Bulletin No. 04,
Software Update ARC_1302 available
Extranet
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced ReaderChapter 3.8 / Page 8 of 13
Type 5151 / 100 / 110 / 200 / 210
Software Releases and Patches
Agfa HealthCare Library*
USB flash drive, delivered with the digitizer
Solution Structure Data Document, Document ID*
30781759
(Intranet Link
CR/DR Interoperability Matrix, Document ID*
31333326
(Intranet Link
Link)
Repair and Service
/
Extranet
Link)
/
Extranet
Link)
(Intranet Link
Agfa Company Confidential
/

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Cr 10-xCr readerCr 12-xCr advanced reader

Table of Contents