DD+DIS155.11E
5.1
Image Plate Run Details
This section describes following details of the image plate run:
#
Description
1
Clamping, unlocking and opening the Cassette
2
Taking the Image Plate out of the Cassette
3
Scanning the Image Plate
5.1.1
Clamping, Unlocking and Opening the Cassette
The user activates following mechanical actions by user inserting the cassette with a
minimum required force of 40 N:
The cassette clamper clamps the cassette.
The cassette unlocker opens the lock of the cassette.
The cassette opener opens the shutter of the cassette.
By pulling out the cassette the opposite actions are activated. In addition the
cassette pusher supports closing the cassette.
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
Lock
Shutter
Cassette Unlocker
Cassette Pusher
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 2 / Page 15 of 19
Type 5151 / 100 / 110 / 200 / 210
Functional Description
Cassette
Cassette Opener
Figure 12
Agfa Company Confidential
See Section
5.1.1
5.1.2
5.1.3
Cassette Unit
Cassette Clamper
517502eb.cdr