(10) Remove the cassette from the digitizer.
(11) Select menu <Logout> and follow the logout wizard.
(12) Close the browser.
2.4

Verifying a successful Shading Calibration

(1)
Expose a flatfield. Use the same exposure parameters like for the
Verification
shading calibration.
(2)
Identify and scan the image. Use exam type: System diagnosis 
Flatfield
IMPORTANT:
By selection of examination type System diagnosis  Flatfield the correct
settings for image processing of the flatfield are used.
Any other examination type may give a result which cannot be used to evaluate the
flatfield image.
(3)
Visually check the image for homogeneity in IP run direction and vertical to the
IP run direction.
Evaluation:
If the scanned flatfield image does not show any lines or stripes or large area
inhomogeneities no further action is required.
If there are unacceptable effects, see service manual chapter 3.3,
Troubleshooting.
The shading calibration is finished.
Result
The controlled version of this document is available from the Agfa HealthCare Library. Any printed copy of this document is uncontrolled.
Edition 1, Revision 1
03-2013
DOCUMENT CONTROL NOTE:
CR 10-X / CR Reader / CR 12-X / CR Advanced Reader Chapter 3.6 / Page 9 of 10
Type 5151 / 100 / 110 / 200 / 210
Repair and Service
Adjustments and Calibrations
Agfa Company Confidential