Dionex Easion Operator's Manual
A.8 Conductivity Detector and Flow Cell
Type
Temperature
Compensation
Cell Drive
Auto Offset
Linearity
Resolution
Calibration
Control and
Data Evaluation
Cell Body
Cell Electrodes
Cell Active
Volume
Maximum Cell
Operating
Pressure
Cell
Temperature
Cell Chemical
Compatibility
88
Bipolar-heated conductivity cell; microprocessor-controlled digital
signal processing over a range of 0 to 10,000 S/cm
Preset for accurate reading at 40 ºC
8 kHz square wave
-9,999 to 9,999 S
1%
0.0047 nS/cm
Factory-calibrated before shipment to the user; calibrated in the
field with memory retention and transfer
Provided by Chromeleon software; communication with the
Dionex Easion is via USB (Universal Serial Bus)
PEEK
Passivated 316 stainless steel
1 L
10 MPa (1500 psi)
Regulated at 40 ºC
Fully compatible with all typical ion chromatography eluents,
including MSA
Doc. 155028-01 08/20