HEIDENHAIN MSE 1000 Operating Instructions Manual page 96

Table of Contents

Advertisement

8
Commissioning and preparing to measure
8.2
Module setup
Teach resolution for a single sensor
To teach the resolution for a single sensor:
Use a gage block to position the sensor so that the Sensor output (V) value
shows as close to 0.0 as possible.
Move the sensor to the minimum mechanical position
Click inside the Min (mm) or Min (in) field and enter the distance that was
traveled from the 0.0 position.
Click the OK button to the right of the Min (mm) or Min (in) field.
Use a gage block to position the sensor so that the Sensor output (V) value
shows as close to 0.0 as possible.
Move the sensor to the maximum mechanical position.
Enter the distance that was traveled from the 0.0 position in the Max (mm) or
Max (in) field.
Click the OK button to the right of the Max (mm) or Max (in) field.
The resolution that will be used to convert from volts to millimeters or inches will
be displayed in the Res. (mm/V) or Res. (in/V) field.
The Calibration timestamp is updated.
Teach resolution for all sensors
To teach the resolution for all sensors:
Use gage blocks to position all of the sensors so that the Sensor output (V)
values show as close to 0.0 as possible.
Move all of the sensors to their minimum mechanical positions.
Click inside each Min (mm) or Min (in) field and enter the distance that was
traveled from the 0.0 position.
Click the SetAllMin button.
Move all of the sensors to their maximum mechanical positions.
Click inside each Max (mm) or Max (in) field and enter the distance that was
traveled from the 0.0 position.
Click the SetAllMax button.
The resolutions that will be used to convert from volts to millimeters or inches will
be displayed in the Res. (mm/V) or Res. (in/V) fields.
The Calibration timestamps are updated.
96
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015