11

Maintenance

A risk of personal injury or death exists if regular inspection and maintenance
tasks are not completed.
Follow the recommended inspection and maintenance schedule to keep the
product in a safe operating condition.
The following inspection and maintenance procedures are required to ensure the
product is maintained in a safe operating state.
Type
Part
Visual in-
Power cord
spection
Visual in-
Symbols and
spection
labels on
product
Visual in-
Product hous-
spection
ing and inter-
face connec-
tions
Electrical
Protective
Test
ground con-
nection
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
WARNING
Interval
Possible failure
Annual
Damaged insula-
tion, visible or dam-
aged wires
Annual
Labels and safety
symbols not read-
able or not present
on the product. Re-
fer to "Safety sym-
bols", page 31.
Annual
Damage or wear
that can affect the
functionality and
safety of the prod-
uct
Annual
Interrupted or bad
connection
Follow up
Replace power
cord
Contact a
HEIDENHAIN ser-
vice technician
Contact a
HEIDENHAIN ser-
vice technician
Replace power
cord or contact a
HEIDENHAIN ser-
vice technician
11