10.1.10 Power-up the modules

Power-up
Press the On (supply) side of the power switch to power-up the system, see
"Initial power-up", page 48.

10.1.11 Configure the network with DHCP

The power LED's on each module light solid green. Network LED's initially flash
green five times per second when waiting for DHCP to assign an IP address.
Wait for the Network LED to flash green twice per second on all modules.
Open the Configuration screen
Click the Config button in MSEsetup.
Enable Supervisor mode
Click the Supervisor mode button.
Enter "95148" into the Password field.
Open the Connect screen
Click the Connect button.
Perform a Broadcast
Select the 255.255.255.0 netmask from the Broadcast netmask drop-down list.
Click the Broadcast button.
Click the Cancel button to proceed without overwriting the settings.
A message verifying that broadcasting is finished is displayed in the Connect
screen message area.
Switch to Static addressing
Click the OK button in the dialog box when asked to switch to static addressing.
10.1.12 Setup the 1 V
Open the Setup screen
From the Connect screen, click the Back button twice.
Click the Setup button.
Select the 1 V
Click on X1 (CH1) under MSE 1184 in the Tree.
Click the Channel tab.
Assign the length gauge to the 1 V
Click and drag the Length Gauge icon to the green square labeled X1. The green
square appears when the length gauge icon is clicked and dragged.
Click the OK button.