Connecting A Transducer; Connecting An I/O Terminal Block Connector - HEIDENHAIN MSE 1000 Operating Instructions Manual

Table of Contents

Advertisement

5.2.7

Connecting a transducer

20
Transducer connector
This product can be used with select Solartron, Tesa, Marposs and Mahr half-
bridge and LVDT transducers. Refer to the transducer documentation for maximum
connecting cable length.
To connect a transducer:
Verify power is removed from all power supply modules in the module chain.
Align the transducer cable connector notch with the module transducer
connection notch
Insert the cable connector into the transducer connection
cable connector by turning clockwise until it is snug
5.2.8

Connecting an I/O terminal block connector

21
I/O terminal block connector
The MSE 1401 (ID 747507-01) provides a terminal block I/O connection for
switching inputs and relay outputs. The switching inputs are active when a
High signal (contact or pulse) is present. They are isolated and can be supplied
externally or internally.
Further Information:
Further Information:
To connect the terminal block:
Verify that power is off or disconnected.
Remove the male terminal connector from the module connection.
Insert the required + and - wires for switching inputs or the required output
wire for relay outputs into their respective terminals on the male terminal
connector.
Using a small flat-edge screwdriver, tighten the screws located on the top of
the cable connector.
Insert the male terminal connector into the connection on the module.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
"Relay outputs", page 179.
"Switching inputs", page 179.
Data interface connections
20
and tighten the