HEIDENHAIN MSE 1000 Operating Instructions Manual page 104

Table of Contents

Advertisement

8
Commissioning and preparing to measure
8.4
Device setup
Enable/disable error monitoring
Error monitoring checks for missing counts and other errors internal to a module.
Diagnostic screen information will be updated when error monitoring is enabled.
Disabling error monitoring is useful for increasing communication speed with
EnDat v2.2 encoders. Function reserves, errors, and warnings will not be
monitored when disabled.
To enable error monitoring:
Click the Error monitoring drop-down arrow.
Select On from the drop-down list.
Click the OK button.
A message verifying the action is displayed in the message area.
To disable error monitoring:
Click the Error monitoring drop-down arrow.
Select Off from the drop-down list.
Click the OK button.
A message verifying the action is displayed in the message area.
Select reference mark
The reference mark setting is used with 1 V
are used to re-establish datum positions after a power interruption.
The following table represents a partial list of HEIDENHAIN encoders. This table
describes the reference mark parameters that must set for the encoders. Most
entries can be found in the operating instructions for the encoder.
To select a reference mark:
Click the Reference mark drop-down arrow and select the required value.
Click the OK button.
A message verifying the action is displayed in the message area.
104
and TTL encoders. Reference marks
PP
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015