Configure the power supply module
Select the power supply module from the Individual Module IP drop-down list.
Click inside the enabled Set Static text field and enter "4" (172.31.46.4).
Click the Set Static button.
MSEsetup sets the static IP address and performs a reconnect to the modules.
Configure the base module
Select the base module from the Individual Module IP drop-down list.
Click inside the enabled Set Static text field and enter "5" (172.31.46.5).
Click the Set Static button.
MSEsetup sets the static IP address and performs a reconnect to the modules.
Set the module chain to use Static addressing
Click the Use Static Addressing button.
A dialog box confirms that all modules have been set to static addressing.
Cycle power for the entire module chain by turning the power off and then back
on.
Wait for the Network LED to flash twice per second on all modules.
Click the OK button in the dialog box.
MSEsetup performs a broadcast. A dialog box confirms the number of modules
found.
Click the OK button in the dialog box to overwrite the configurable module
settings.
10.2.9

Power-off the modules

Power-off
Press the Off (supply) side of the power switch to power-off the system.

10.2.10 Mount the EnDat module

Mount EnDat module
Mount the MSE 1314 (ID 747503-01) 4-channel EnDat module, see "Mounting a
module", page 34.
Connect base and EnDat module
Connect the base module and the 4-channel EnDat module, see "Connecting
modules", page 35.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Example using manual IP address assignment
10