Commissioning and preparing to measure
8.4
Device setup
Assign line count
The line count is based on the type of 1 V
count is used to convert the count returned from the encoder to a rotary position.
Refer to the documentation provided with the device for the correct line count
setting.
To assign a line count:
Click inside the Line count text field.
Delete the current value.
Enter the required line count value.
Click the OK button.
A message verifying the action is displayed in the message area.
Assign interpolation
The interpolation setting is used with TTL encoders. The interpolation is used
in conjunction with the signal period or line count when converting the count
returned from the encoder to a position.
To assign the interpolation:
Click the Interpolation drop-down arrow and select the required value.
Click the OK button.
A message verifying the action is displayed in the message area.
Teach gain
Teaching gain is used with LVDT sensors. Teaching gain sends a command to the
LVDT module to determine the optimal gain value. The gain value is used to select
the best voltage range for the sensor.
To teach the gain:
Click the OK button to the right of the Teach gain text field.
A dialog window appears requesting to hold the sensor at the minimum position.
Hold the sensor at the minimum position and click the OK button.
A dialog window appears requesting to hold the sensor at the maximum position.
Hold the sensor at the maximum position and click the OK button.
A dialog window appears requesting to hold the sensor at either the minimum or
the maximum position.
Hold the sensor at the minimum or maximum position and click the OK button.
A dialog window appears informing if the teach gain passed or failed.
The Gain code will be updated with the new value.