Commissioning And Preparing To Measure; Network Configuration; Connect Screen - HEIDENHAIN MSE 1000 Operating Instructions Manual

Table of Contents

Advertisement

8

Commissioning and preparing to measure

8.1

Network configuration

MSEsetup communicates with MSE 1000 modules over Ethernet. The PC running
MSEsetup is referred to as the Workstation and is the Client when communicating
with the modules.
The workstation must have a Network Interface Controller (NIC) with a unique
Internet Protocol (IP) address that is on the same subnetwork as the MSE 1000
modules. All MSE 1000 modules must also have a unique IP address.
Default IP addresses for similar module types are set the same from the factory
and need to be reconfigured in order to communicate with MSEsetup. Workstation
and MSE 1000 module IP address configurations and other network settings and
operations are located in the Connect screen.
Model
MSE 120x
MSE 11xx, MSE 13xx, MSE 1401,
MSE 1501
8.1.1

Connect screen

The Connect screen is available without a password when MSEsetup is opened
for the first time. All subsequent MSEsetup sessions require enabling Supervisor
mode to access the Connect screen.
Available options:
Workstation IP
IP address
Netmask
Port
Async. Port
Connection
Disconnect
Reconnect
Broadcast
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Network configuration
Default IP ad-
Default netmask
dress
172.31.46.2
255.255.255.0
172.31.46.1
Individual Module IP
Query
Set Static
Module chain
Use DHCP Addressing
Use Static Addressing
Restore Factory Defaults
8
8.1
77