8.1.8

Configuring the module chain manually

The MSE 1000 module chain must be configured manually when a DHCP enabled
router is not used.
An RJ-45 crossover cable may be required when connecting the
workstation directly to the base module. Refer to the documentation
provided with the NIC for additional information.
Configure the power supply and base module
The power supply and base module must be configured first when configuring the
module chain manually.
To configure the power supply and base module manually:
Install the first power supply module and the base module.
Turn the power on.
The power LED's on each module should light solid green. Network LED's initially
flash green five times per second when waiting for DHCP to assign an IP address.
After a 45 second timeout the modules will stop waiting for DHCP to assign an IP
address and the Network LED's on each module will flash green twice per second.
Wait for the Network LED to flash twice per second on all modules.
Select the power supply module from the Individual Module IP drop-down list.
Assign the power supply module a unique static IP address, see "Individual
Module IP", page 83.
Assign the base module a unique static IP address.
Configure the module chain to use static IP addresses, see "Module Chain",
page 84.
Verify that the IP address for each module has been changed to the new static
IP address value.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Network configuration
8
8.1
89