Open Function reserves screen
To open the Function reserves screen:
Click the Function reserves button.
The Function reserves screen is displayed in the content area.
LVDT sensor diagnostics
LVDT sensor diagnostics are used to monitor the LVDT sensor output voltage.
The LVDT sensor diagnostics screen displays a graphical representation of how
much of the total positional range of the electronics is being utilized. The yellow
areas of the graphic signify the outermost 20% of the total usable range. The
positional accuracy and stability of the sensor is at its best when closest to the
center of the graphic. The center of the LVDT sensor is also known as the NULL
position.
LVDT sensor diagnostics:
Sensor output
LVDT sensor diagnostics screen
LVDT sensor diagnostics
Sensor output (V)
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Displays the voltage output of the sensor.
Diagnostics
12.1
147
12