Module Configuration File Errors; Device Warnings And Errors; Encoder Warnings And Errors - HEIDENHAIN MSE 1000 Operating Instructions Manual

Table of Contents

Advertisement

12
Troubleshooting
12.8

Encoder warnings and errors

12.6

Module configuration file errors

Error
Invalid file
DOM root element is NULL
Invalid tagname for Module-
Config
Invalid tagname for the module
Invalid tagname for the chan-
nel
Invalid module list
Invalid channel list
Tag name not found
Failed to overwrite all of the se-
lected modules configurable
settings
12.7

Device warnings and errors

Warning
Calibration needs to be per-
formed
Mastering must be redone
since the device settings
changed
12.8
Encoder warnings and errors
Encoder warnings and errors are reported in the Error and Diagnostic screens.
Further Information:
Further Information:
Refer to the documentation provided with the encoder for troubleshooting information.
158
Cause
The requested configuration file
is invalid.
The root element of the config-
uration file is not valid.
The ModuleConfig tag name
was not found in the configura-
tion file.
The Module tag name was not
found in the configuration file.
The Channel tag name was not
found in the configuration file.
The tag name for the desired
module element could not be
found.
The tag name for the desired
channel element could not be
found.
The tag name for the desired
element could not be found.
The file being merged with the
ModuleConfig.xml file is invalid.
Cause
An analog or LVDT device
needs to be calibrated.
A device's setting has been
changed that has invalidated
the master position.
"Error screen", page 149.
"Diagnostics", page 140.
Corrective action
Load a good Module
Configuration file, see "File
options", page 71.
Restart MSEsetup.
or
Delete the Module
Configuration file, see
"Installing MSEsetup",
page 49 for configuration
file locations.
Perform a broadcast.
Choose a different
file to merge with the
ModuleConfig.xml file.
Corrective action
Recalculate the resolution of
the analog or LVDT device.
Clear the warning and re-
master the device from the
master screen.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015