HEIDENHAIN MSE 1000 Operating Instructions Manual page 107

Table of Contents

Advertisement

Set gain code
Setting the gain code is used with LVDT sensors. The gain value should be taught
first and setting the gain should be used if more or less resolution is desired. The
position may not be as stable when too much resolution is used. Setting the gain
to a value that is too small may not give enough resolution.
To set the gain code :
Click inside the Gain code text field.
Delete the current value.
Enter the new value.
Click the OK button.
A dialog window appears requesting to hold the sensor at the min position.
Click the Cancel button if only the gain code needs to be set, otherwise click
the OK button and set the minimum and maximum positions.
Set measured minimum or maximum
The measured minimum and measured maximum value is used with analog
devices and LVDT sensors. The calculation of the resolution for analog devices
will be done by interpolating the signal minimum and signal maximum over the
measured minimum and measured maximum. The calculation of the resolution
for LVDT sensors will be done by prompting the user to move the sensor to the
mechanical minimum and maximum before interpolating.
To set the measured minimum and maximum values for analog devices:
Click inside the Measured min. text field.
Delete the current value.
Enter the minimum value.
Click inside the Measured max. text field.
Delete the current value.
Enter the maximum value.
Click the OK button.
The resolution will be updated with the interpolated value.
The calibration time-stamp will be updated with the current date and time.
A message verifying the action is displayed in the message area.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Device setup
8
8.4
107