HEIDENHAIN MSE 1000 Operating Instructions Manual page 108

Table of Contents

Advertisement

8
Commissioning and preparing to measure
8.4
Device setup
To set the measured minimum and maximum values for LVDT sensors:
Click inside the Measured min. text field.
Delete the current value.
Enter the minimum value.
Click inside the Measured max. text field.
Delete the current value.
Enter the maximum value.
Click the OK button.
A dialog window appears requesting to hold the sensor at the minimum position.
Hold the sensor at the minimum position and click the OK button.
A dialog window appears requesting to hold the sensor at the maximum position.
Hold the sensor at the maximum position and click the OK button.
The resolution will be updated with the interpolated value.
The calibration timestamp will be updated with the current date and time.
A message verifying the action is displayed in the message area.
Set recalibration timer
The recalibration timer is used with analog devices and LVDT sensors. The
recalibration timer provides a notification in the Error screen when it is time to
recalibrate the resolution of a device or sensor.
To set the recalibration timer:
Click inside the Recal. timer text field.
Delete the current value.
Enter the required recalibration timer value.
Click the OK button.
A message verifying the action is displayed in the message area.
108
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015