8
Commissioning and preparing to measure
8.1
Network configuration
Configure additional modules
MSE 1000 modules must be configured one at a time when configuring the
module chain manually.
To configure additional modules:
Turn the power off.
Install the next module into the module chain.
Turn the power on.
Wait for the Network LED to flash twice per second on all modules.
Perform a broadcast, see "Connection", page 81.
If the new module is configured to use DHCP addressing (default) a dialog box will
appear asking to configure the new module to use static addressing.
Click the OK button to switch to static addressing.
A dialog box confirms the module has been configured to use static addressing.
Cycle the power. Turn the power off, wait 20 seconds then turn the power on.
Wait for the Network LED to flash twice per second on all modules.
Click the OK button in the dialog box.
A dialog box displays the number of modules found.
Click the OK button to overwrite the configurable module settings from a
backup file or click the Cancel button to proceed without overwriting the
settings.
Assign the new module a unique static IP address.
Repeat this process until all required modules are configured into the module
chain.