10.2.15 Configure the EnDat module network communication settings

The power LED's on each module light solid green. Network LED's flash green
twice per second.
Wait for the Network LED's to flash green twice per second on all modules.
Perform a Broadcast
Click the Broadcast button.
A dialog box appears asking to configure the EnDat module to use static
addressing.
Configure the EnDat module for static addressing
Click the OK button.
A dialog box confirms the EnDat module has been configured to use static
addressing.
Cycle the power
Turn the power off, wait 20 seconds then turn the power on.
Wait for the Network LED's to flash twice per second on all modules.
Click the OK button in the dialog box.
Click the OK button to overwrite the settings.
Configure the EnDat module IP address
Select the EnDat module from the Individual Module IP drop-down list.
Click inside the enabled Set Static text field and enter "6" (172.31.46.6).
Click the Set Static button.
MSEsetup sets the static IP address and performs a reconnect to the modules.
10.2.16 Setup the 1 V
Open the Setup screen
From the Connect screen, click the Back button twice.
Click the Setup button.
Select the 1 V
Click on X1 (CH1) under MSE 1184 in the Tree.
Click the Channel tab.
Assign the length gauge to the 1 V
Click and drag the Length Gauge icon to the green square labeled X1. The green
square appears when the length gauge icon is clicked and dragged.
Click the OK button.
Further Information:
options.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Example using manual IP address assignment
channel
PP
Channel
PP
channel
PP
"Channel setup", page 97 for additional channel settings
10
10.2
135