10.2.4
Install the software
Install MSEsetup
Download and install the MSEsetup application software, see "Installing
MSEsetup", page 49.
10.2.5
Open the software
Open MSEsetup
Double-click the MSEsetup desktop icon to open MSEsetup, see "Basic
functions", page 61.
10.2.6
Configure the workstation NIC
Configure the workstation IP
Configure the workstation NIC IP address to 172.31.46.3 in the workstation PC
networking settings. Refer to the documentation provided with the PC or NIC
for configuration instructions.
Configure the workstation Subnet mask
Configure the workstation Subnet mask to 255.255.255.0 in the workstation PC
networking settings. Refer to the documentation provided with the PC or NIC
for configuration instructions.
10.2.7
Power-up the modules
Power-up
Press the On (supply) side of the power switch to power-up the system, see
"Initial power-up", page 48.
10.2.8
Configure the workstation IP , power supply, and base module
network communication settings
The power LED's on each module light solid green. Network LED's initially flash
green five times per second when waiting for DHCP to assign an IP address.
After a 45 second timeout the modules will stop waiting for DHCP to assign an IP
address and the Network LED's on each module will flash green twice per second.
Wait for the Network LED to flash green twice per second on all modules.
HEIDENHAIN | MSE 1000 | Operating Instructions | 10/2015
Example using manual IP address assignment
10
10.2
131