Front End Cooling System

5-8-1

General Description

The Vivid™ 3 Front End cooling system includes following components:
Dust Filter
Fans
Other components that have an influence on the airflow is the Front End itself. Air is sucked from the
inside of the system through the Filter Cover and the Filter. It passes the Fan Assembly, directly into the
upper part of the Front End where it forms a cooling air current between the Boards, before being blown
out against the floor. See
5-8-2

Location in the Unit

The Fan Assembly is located on top of the Front End Crate.