Board Rack Cooling System

5-14-1

General Description

The Board Rack cooling system includes following components:
Dust Filter
Dust Filter Frame
Air Inlet channel
Fan Assembly
Other components that influence on the airflow is the Board Rack itself. The air is sucked in from the
rear side of the system through the Filter Cover and the Filter. It passes the Fan Assembly directly into
the upper part of the rear side of the Board Rack and makes a cooling air current between the Boards
before it blows out against the floor. See Figure 5-50.
5-14-2

Location in the Unit

The Fan Assy is located between the Board Rack and the Filter, which can be seen on the rear side of
the System.
Airflow through Board Rack
behind Side Cover
Air outlet under the System
5 - 96
02
EVISION
Figure 5-50 Airflow through the System
Section 5-14 - Board Rack Cooling System
V
7 / V
7 PRO S
IVID
IVID
ERVICE
Air inlet through Filter
Fan Assy
M
ANUAL