Mechanical Functions Checks; Kickstand Mechanical Checks - GE Venue Fit Technical Publication

Table of Contents

Advertisement

P R E L I M I N A R Y
DIRECTION 5854713-100, REVISION 2
VENUE FIT™ SERVICE MANUAL
4-3-6

Mechanical Functions Checks

4-3-6-1

Kickstand mechanical Checks

1.) Lay the system on a stable flat surface covered with a soft material (such as foam or thick cloth) to
prevent scratches to the display.
2.) Pull upwards the kickstand, ensuring that the kickstand doesn't move freely
Figure 4-57 Kickstand
3.) Carefully position the system, leaning on the kickstand.
4.) Verify that the system and the kickstand are stable and do not "slip".
Chapter 4 - General Procedures and Functional Checks
4-21