Download Print this page

Agilent Technologies 2200 TapeStation Quick Manual

Screentape system
Hide thumbs Also See for 2200 TapeStation:

Advertisement

Quick Links

The Agilent 2200 TapeStation system is an automated platform for simpler, faster and more reliable
electrophoresis. It is made up of 3 elements:
• 2200 TapeStation system (G2964AA) or 2200 TapeStation Nucleic Acid system (G2965AA)
• High Sensitivity D1000 ScreenTape (5067- 5584) with High Sensitivity D1000 Reagents (5067- 5585)
• Agilent Software packages (2200 TapeStation Controller Software, and TapeStation Analysis Software)
Kits
The High Sensitivity D1000 ScreenTape system is designed for analysing DNA molecules from 35 – 1000 bp.
Specifications
Analytical Specifications
Sizing Range
Typical Resolution
1
Sensitivity
Sizing Precision
2
Sizing Accuracy
Quantitative Precision
3
Quantitative Accuracy
Quantitative Range
Maximum sample buffer strength
Physical Specifications
Analysis Time
Samples per consumable
Sample volume required
Kit stability
Kit size
1
signal- to- noise >3 (single peak)
2
Sizing Accuracy for electronic ladder: ±20%
3
Measured against 2100 Bioanalyzer system
Agilent High Sensitivity D1000
ScreenTape System Quick Guide
High Sensitivity D1000 ScreenTape assay and reagents
35 – 1000 bp
35 – 300 bp: 15 %; 300 – 1000 bp: 10 %
5 pg/µL
5 % CV
±10 %
15 % CV
±20 %
10 – 1000 pg/µL
7 mM KCl, 20 mM Phosphate Buffer, 20 mM Guanidine-HCl, 80 mM NaCl, 20 mM Acetate
16 samples: <20 min; 96 samples: ≈100 min
16
2 µL
4 months
112 samples
Agilent Technologies

Advertisement

loading

Summary of Contents for Agilent Technologies 2200 TapeStation

  • Page 1 Agilent High Sensitivity D1000 ScreenTape System Quick Guide The Agilent 2200 TapeStation system is an automated platform for simpler, faster and more reliable electrophoresis. It is made up of 3 elements: • 2200 TapeStation system (G2964AA) or 2200 TapeStation Nucleic Acid system (G2965AA) •...
  • Page 2 Additional Material Required (Not Supplied) • Volumetric pipette • Centrifuge 2200 TapeStation instruments are supplied with an optional IKA MS3 vortexer which N O T E includes a 96-well plate adaptor suitable for both 96-well PCR plates and 8-way strips.
  • Page 3 C A U T I O N ➔ Use only the recommended consumables and reagents with the 2200 TapeStation system. General Information on Working with DNA • For best results ensure that all reagents are allowed to equilibrate to room temperature N O T E for 30 min prior to use.
  • Page 4 Parts required Description 5067-5584 High Sensitivity D1000 ScreenTape Launch the 2200 TapeStation Controller Software. Load High Sensitivity D1000 ScreenTape device and loading tips into the 2200 TapeStation instrument. Sample Preparation High Sensitivity D1000 ScreenTape Assay Parts required Description 5067-5585 High Sensitivity D1000 Reagents...