Timing Trigger Macro Library - Agilent Technologies 16517A User Reference

4-gsa/s timing and 1-gsa/s synchronous state logic analyzer
Table of Contents

Advertisement

The Trigger Menu

Timing Trigger Macro Library

Timing Trigger Macro Library
The following list contains all the macros in the Timing Trigger Macro
Library. They are listed in the same order as they appear onscreen.
User Mode
User level - custom combinations, loops
The User level is a user-definable level. This level offers low level
configuration and uses one internal sequence level.
Basic Macros
1. Find anystate n times
This macro becomes true when the first state it sees appears "n" times. It
uses one internal sequence level.
2. Find pattern present/absent for > duration
This macro becomes true when it finds a designated pattern that has been
present or absent for greater than or equal to the set duration. It uses one
internal sequence level.
3. Find pattern present/absent for < duration
This macro becomes true when it finds a designated pattern that has been
present or absent for less than the set duration. It uses one internal
sequence level.
4. Find edge
This macro becomes true when the designated edge is seen. It uses one
internal sequence level.
5. Find Nth occurrence of pattern present/absent for > duration
This macro becomes true when it finds the Nth occurrence of a pattern that
has been present or absent for greater than or equal to the set duration.
Each valid pattern occurrence is only counted once, even though the same
occurrence may be valid for multiple sample clocks. It uses one internal
sequence level.
4–12

Advertisement

Table of Contents
loading

This manual is also suitable for:

16518a

Table of Contents