Summary of Contents for Thermo Scientific F1-ClipTip
Page 1
Thermo Scientific F1-ClipTip Single Channel Multichannel Instructions for Use...
Page 2
The CE mark covers the system containing the CE marked F1-ClipTip pipette and the CE marked ClipTip pipette tips. Product specifications are subject to change without prior notice. ClipTip is a registered...
0.1-2 µl and 10 µl based pipette models use ‘Snap and Lip’ seal tips. The F1-ClipTip product offering covers a volume range from 0.1 µl to 1000 µl and includes variable and fixed volume single channel pipettes as well as multichannel pipettes. The convenient color coding* system facilitates to identify the correct ClipTip by volume for use.
2. Package The F1-ClipTip package contains the following items: F1-ClipTip pipette Service tool Multichannel service tool Tip fitting tool Bag of grease 1g (Order No. 3300200) Instructions for Use Calibration certificate / Warranty Certificate O-ring Order No. 2214930 Single Channel Pipette (Scp) 300 (3 pcs) Order No.
The F1-ClipTip is to be used by trained personnel with required laboratory skills. The instructions for use must be read prior to and during the use of the device (pipette and tip). The F1-ClipTip pipette can be used between +4°C and +40°C.
Operate the pipetting button gently. If it is quickly released, sample liquid may enter the pipette, which will affect the pipetting performance and may damage the pipette. Follow the guidance in the “Thermo Scientific Good Laboratory Pipetting Practices” brochure to minimize the risk of repetitive strain injury (RSI).
Page 8
2. Lift the pipette. 3. If the tip is not attached, repeat Steps 1 and 2. Incorrect use of tips Use only ClipTip tips with an F1-ClipTip pipette. Check that all clips are locked to ensure tip sealing. Picture 4b ClipTip tips are designed for single use only.
6. Calibration and Adjustment 6.1 Factory calibration limits F1-ClipTip pipettes are factory calibrated and adjusted to give the volumes as specified with distilled or deionized water using the forward pipetting technique. Manufacturer specifications are shown on the pipette calibration certificate and in Appendix 2. The pipette should be recalibrated for use with any tip type not named on the Calibration Certificate.
Page 10
The performance of a new and serviced pipette must be checked and documented. The manufacturer’s specifications should be used as guidelines and the user should establish acceptable imprecision and inaccuracy performance limits (ISO 8655). The user determines the performance and applicability of the pipette and tip combination for a particular application and determine the required interval for performance checking.
7.2.1 Disassembly of 1000ul single channel pipettes 1. Remove the tip fitting 5 (Picture 8). To remove the tip fitting place the star-shaped end of tool 3 (picture 9) into the end of the tip fitting. Turn the tool counterclockwise to unscrew the tip fitting (picture 10). Picture 9 Picture 8 2.
3. Remove the Clip trigger 8 and Clip trigger spring 9. 4. Turn the tip cone 11 counterclockwise with the service tool 1. There are two slots in the service tool: the smaller one for pipettes under 300 μl and the bigger one for pipettes over 300 μl.
Reassemble the Clip trigger spring 9 and Clip trigger 8. Assemble the tip fitting according to instruction in 7.2.4. Reassemble the tip ejector 10. 100-1000μl, 500 / 1000μl Fixed Put the O-ring 17 and support ring 16 onto the tip cone 11. Slide the spring 13 onto the piston 12 and slide the entire assembly into the tip cone 11.
Page 15
7. While the tip ejector 23 is in the down position, open its upper end slightly on both sides and pull downwards. Remove the fastening spring 22. Remove the ejector link 24. Press the piston bar 16 beam into the down position. Finally unfasten the four screws 20/21 and the tip ejector. The ejector slide 25 and two springs 26 can be removed for maintenance.
7.2.7 Cleaning and reassembling multichannel pipettes See the cleaning instructions in Section 7.2.2 “Single channel pipettes”. Check the volume-specific reassembling of the piston assemblies starting on page 23. The pipette is assembled in the reverse order to disassembling. Note that the ejector link 24 must be in the down position simultaneously with the tip ejector 23 (page 26).
9. Troubleshooting The table below lists possible problems and their solutions. Defect Possible reason Possible action ►Re-attach/tighten the tip fitting Leakage • Tip fitting incorrectly attached or loose or replace with new tip fitting(s) using the tool. ►Discard the tip. • ClipTip clips are bent ►Change the O-ring.
10. Appendices Appendix 1. Conversion table Value of the conversion factor Z (µl/mg), as a function of temperature and pressure, for distilled water. Temperature Air pressure °C 101.3 15.0 1.0017 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020 15.5 1.0018 1.0019 1.0019 1.0020 1.0020 1.0020...
Page 21
Appendix 3. ISO8655 calibration specifications Thermo Scientific F1-ClipTip specifications according to ISO8655 standard. Thermo Scientific F1-ClipTip Variable Volume Single Channel Range Volume Inaccuracy Imprecision Compatible Color µl µl SD µl ClipTip code 0.1–2 µl ±0.080 ±4.0 0.040 12.5 Ext pink ±0.080...
Page 22
Appendix 4. Assembly and spare parts * Available as a spare part. See model specific order numbers on page 30. 1062930 1062800 Multichannel service tool Service and adjustment tool 0300070 Tip fitting tool * 5. Tip fitting * 42. Tip cone assembly * 50.
Need help?
Do you have a question about the F1-ClipTip and is the answer not in the manual?
Questions and answers