Heat; View Total Operating Hours - Thermo Scientific MaxQ SHKE8000 Series Operating And Maintenance Manual

Stackable orbital shakers
Hide thumbs Also See for MaxQ SHKE8000 Series:
Table of Contents

Advertisement

Section 2
Operation
View Total Operating
2-18
MaxQ SHKE8000 Series Shakers
To view this information, open the Configuration menu by pressing the
down arrow, up arrow, and the Silence button, and then the down arrow
Hours
three times, in the sequence shown below.
The screen at the right will appear
on the display.
Pressing the Time button beneath
RunHrs shows the total
accumulated run hours as
displayed at right. When finished,
press any of the three buttons
(Time, Speed, or Temp) to save
the new setpoint value and return to the previous screen, or press nothing
for about 15 seconds to save the new setpoint value and return to the
Operating Screen.

Heat %

Heat percentages are intended for factory use only, and can be helpful in
troubleshooting the heat control system.
To view this information, open the Configuration menu by pressing the
down arrow, up arrow, and the Silence button, and then the down arrow
three times, in the sequence shown below.
The screen shown at the right
will appear on the display.
Press the Temp button beneath Heat %.
Actual
Total Run Time:
Setpoints
Hours
Thermo Scientific
XX

Advertisement

Table of Contents
loading

This manual is also suitable for:

Shke8000Shke8000-1ceShke8000-7Shke8000-8ce

Table of Contents