Chapter 1
Laboratory Preparation
CAUTION
Do not directly expose the GC system to any cooling duct outlets.
The air conditioning load for a basic FOCUS GC system is approximately 400 W
(3866 Btu h
FOCUS GC standard system and FOCUS GC optional instruments.
NOTE
Other peripheral devices, such as purge and trap units, may be connected to the
FOCUS GC . These peripheral devices may have additional cooling requirements. Consult
your local Thermo Fisher Scientific CSE for assistance.
28
-1
). Table 1-5 and Table 1-6 show the approximate heat output for a
Table 1-5. Heat Output: FOCUS GC Standard System
Instrument
FOCUS GC
computer
monitor
Total
Table 1-6. Heat Output: FOCUS GC Optional Instruments
Instrument
GC Autosamplers
AI/AS 3000
TriPlus
Printers
ink jet printer/plotter
laser printer
Operating Environment
Heat Output
(in BTU h
1366
1650
850
3866
Heat Output
(in BTU h
341.5
752
200
2900
Site Preparation and Installation Manual