4 Maintenance Procedures

Fitting the extraction cone to the ion block

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the ion block.
To fit the extraction cone to the ion block:
1.
Fit the extraction cone handle insulator to the extraction cone handle.
2.
Fit the extraction cone to the ion block.
3.
Rotate the 2 PEEK extraction cone retainer clips to secure the extraction
cone, then use the combined 2.5-mm Allen wrench and cone extraction
tool to tighten the retainer clip securing screws.

Fitting the ion block assembly to the source assembly

Required materials
Chemical-resistant, powder-free gloves
Combined 2.5-mm Allen wrench and cone extraction tool
Warning:
toxic materials, and to avoid spreading contamination to
uncontaminated surfaces, wear clean, chemical-resistant,
powder-free gloves when working with the source components.
Warning:
working with the source enclosure open if one or both of these
conditions apply:
• An ESI probe is fitted (the probe tip is sharp).
• A corona pin is fitted (the pin tip is sharp).
126
February 11, 2013, 71500130002 Rev. D
To avoid personal contamination with biohazards or
To avoid personal contamination with biohazards or
To avoid puncture wounds, take great care while