LCD Flex Cable Replacement Procedure

8-3-9-1
Tools
Use the appropriate Phillips and flat screw drivers, as indicated in the LCD Flex Cable replacement
procedure.
8-3-9-2
Time Required
15 minutes
8-3-9-3
Preparation
Shut down the Vivid i n/ Vivid q N ultrasound unit, as described in - System Setup.
8-3-9-4
LCD Flex Cable Removal Procedure
1) Disconnect the AC cable and battery.
2) Remove the Control Panel and Keyboard as described in the
Procedure on page 8 -
3) Remove the LCD Display Frame as described in the
page 8 -
4) Remove the LCD Display as described in the
5) Using both hands, carefully lift the LCD Display upwards and remove it from the Rear Cover & Latch
Assembly, as shown in
NOTE:
Take care not to damage the LCD Flex ribbon cable when performing this step.
Flex Cable Replacement Procedure
8-58
1
EVISION
9.
27.
Figure
Figure 8-68 Removing the LCD Display
Section 8-3 - Internal Component Replacement Procedures
LCD Display Frame Removal Procedure on
LCD Display Removal Procedure on page 8 -
8-68.
V
V
N S
IVID I N AND
IVID Q
Control Panel and Keyboard Removal
M
ERVICE
ANUAL